Procedure
The source motion board can now be
disconnected from the motion controller.
17. Assemble the new source motion board
and the motion controller. Connect the
motion controller interface cable.
18. Place the motion controller and AC source
motion board assembly in the line source
compartment.
19. Mount the four Allen screws holding the
motion controller and AC source motion
board assembly to the bottom of the line
source compartment.
Remember to include the cable tie holding
the collision cable.
Note. The screw holding the cable tie is longer
than the other screws and must be placed here.
20. Mount the four Allen screws holding the
source motion board.