Repair Procedures
Procedure
7. To ensure that the transportation screws
can always be found when required, place
them in the spare holes provided in the
line source housing.
8. Tighten the transportation screws to
ensure that they will not come loose and
fall out during operation of the CardioMD
AC option.
9. After exchanging the line source, grease
the acme screw inside the line source
compartment. See section 12.8.1, page 12-
116 for instructions.
Returning the Used Line Source to the Vendor
The site Radioactive Materials License holder must return used Gd-153 lines sources to:
Isotope Products Laboratories
Medical Imaging Division
24937 Tibbitts Avenue
Valencia, Ca 91355
When shipping line sources, the site Radioactive Materials License holder should either use the
original packaging material or the packaging material that came with the replacement line sources.
12-74 CardioMD Service Manual
Details
WARNING
Gd-153 source management is the sole responsibility of the site Radioactive
Materials License holder. Philips Medical Systems employees are not
authorized to participate in the legal, safe return of radioactive sources and
should not be involved in the packaging, documentation, and/or shipment of
radioactive sources.
Transportation screws
1MAN0158-H11 04-2005 CSIP Level 1