Evaluation

Calculate the deviation between the measured area dose product and the displayed area-
dose product:
The deviation must be < 0.3 .
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ]
/ (measured area dose product FDPg)
For deviations > 0.3, the area dose product measurement device must be replaced.