6 - 18

VIDEOMED DC

Cut off the cable ties (refer to the arrow, Fig. 6) and disconnect the plugs.
Remove plug X1 at the Compact optics.
Remove the VIDEOMED DC and place it on a clean surface. Refer to: Replacing the
VIDEOMED DC.
1
3
2
Fig. 7

High voltage section of the I.I. mini voltage supply

Remove the high voltage section of the I.I. mini voltage supply. Remove both attach-
ment screws at the C-arm flange. Refer to the arrows, Fig. 6.
Disconnect the anode cable from the high voltage section.
Remove the O-ring and the cap nut from the cable.

Compact optics

NOTE
Remove the attachment screws of the Compact optics. If necessary, remove the retain-
er clip for the plug / cable leads as well.
Remove the Compact optics.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
5
3
2
4
When removing the Compact optics, make sure that no dust or
dirt particles contaminate the I.I. output.
Do not loosen the eccentric screws (2/Fig. 7) on the edge of the
Compact optics! These screws are used to center the Compact
optics to the I.I. output.
SPR2-230.061.01
Rev. 13
Page 18 of 48
12.04
CS PS 24
6
6
6
Siemens AG
Medical Solutions