Cut off the cable ties (refer to the arrow, Fig. 6) and disconnect the plugs.
•
Remove plug X1 at the Compact optics.
•
Remove the VIDEOMED DC and place it on a clean surface. Refer to: Replacing the
VIDEOMED DC.
1
3
2
Fig. 7
High voltage section of the I.I. mini voltage supply
•
Remove the high voltage section of the I.I. mini voltage supply. Remove both attach-
ment screws at the C-arm flange. Refer to the arrows, Fig. 6.
•
Disconnect the anode cable from the high voltage section.
•
Remove the O-ring and the cap nut from the cable.
Compact optics
NOTE
•
Remove the attachment screws of the Compact optics. If necessary, remove the retain-
er clip for the plug / cable leads as well.
•
Remove the Compact optics.
SIREMOBIL Iso-C
System Manual
Replacing boards / replacing components
5
3
2
4
When removing the Compact optics, make sure that no dust or
dirt particles contaminate the I.I. output.
Do not loosen the eccentric screws (2/Fig. 7) on the edge of the
Compact optics! These screws are used to center the Compact
optics to the I.I. output.