Replacing boards / replacing components

Mouse, external mouse and PC keyboard

NOTE
The system comes with a mouse and an external mouse. Both are connected to the
mouse interface of the 3D reconstruction PC by means of a mouse Y-switch.
The keyboard is connected to the keyboard interface of the 3D reconstruction PC.

Replacing the mouse

Switch the system off.
Open the rear panel of the monitor trolley and the paneling of the 3D reconstruction PC.
Replace the mouse. To do so, pull the mouse connector out of the mouse Y-switch, cut
the cable ties at the spring steel band and pull the cable out of the cable duct.
Run the cable of the new mouse through the cable duct and along the spring steel band
and reconnect it.
Reattach the cable from the mouse Y-switch to the keyboard compartment to ensure that
there is sufficient cable length in the keyboard compartment.
Using cable ties, tie the cable (incl. keyboard cable) to the spring steel band again at a
distance of approx. 10 cm.
Perform a function check of the mouse, the external mouse and the keyboard.
Reattach the paneling.

Replacing the external mouse

Switch the system off.
Unplug the external mouse at the rear and replace it with the new external mouse.
Perform a function check of the mouse, the external mouse and the keyboard.

Replacing the PC keyboard

Switch the system off.
Open the rear panel of the monitor trolley and the paneling of the 3D reconstruction PC.
Replace the keyboard. To do so, unplug the keyboard at the 3D reconstruction PC, cut
the cable ties at the spring steel band and pull the cable out of the cable duct.
Run the cable of the new keyboard through the cable duct and along the spring steel
band and reconnect it.
Using cable ties, tie the cable (incl. mouse cable) to the spring steel band again at a dis-
tance of approx. 10 cm.
Perform a function check of the mouse, the external mouse and the keyboard.
Reattach the paneling.