Adjustments / Programs
For the query "Do you want to reset the unit", click on "No". The X-Iris is automatically
closed to the smallest format.

Full format, X-Iris closed

Briefly release fluoro. The X-Iris is displayed on the monitor.
Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-Iris.
In the "Available Functions:" window select "Close Pos - II. Full Format" and click on "Ex-
ecute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click on "No".
The X-Iris remains in the minimum position.
Zoom format is automatically selected.

Zoom format, X-Iris closed

Briefly release fluoro. The X-Iris is displayed on the monitor.
Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-Iris.
In the "Available Functions:" window select "Close Pos - II. Zoom Format" and click on
"Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit" click on "No".
The X-Iris is opened automatically.
Zoom format remains selected.

Zoom format, X-Iris open

Briefly release fluoro. The X-Iris is displayed on the monitor.
Use the keys for opening and closing the X-Iris to set the diameter of the superimposed
circle to the size of the X-Iris.
In the "Available Functions:" window select "Open Pos - II. Zoom Format and click on
"Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click on "No".
Siemens AG
Medical Solutions
SPR2-230.061.01
Rev. 13
12.04
Page 15 of 30
SIREMOBIL Iso-C
CS PS 24
5 - 15
5
5
5
System Manual