Installing Power Cables For Each Instrument - HP 4062UX Installation Manual

Semiconductor process control system semiconductor parametric test system
Table of Contents

Advertisement

5. Remove the PDU from the cabinet.
10. Reinstall the clear plastic cover on the power
cable terminal unit.
11. Fasten the cable clamp securely.
12. Reinstall the top cover of the PDU.
13. Reinstall the PDU in the cabinet.

Installing Power Cables for Each Instrument

Power cables for each rack mounted instrument are installed before the HP 4062 system is
shipped from the factory. Before you turn the HP 4062 on, however, make sure that all power
plugs are completely inserted into the outlets of the PDU.
Before you turn the HP 4062 on, make sure that the line voltage selector
Note
switch of each instrument is set correctly. Refer to the
each instrument for instructions on how to set these switches.
According to the system con guration, you may not be able to supply electric
Note
power from the PDU to other instruments, controllers, and peripherals. See
the
Pre-installation Manual
If the line voltage in your area is between 220 Vac and 240 Vac, connect each
Note
instrument, controller, or peripheral to the PDU via the power cable (HP part
number 8120-1763) when you want to supply the power from the PDU.
6. Remove the top cover of the PDU.
7. Pass the main power cable through the rear
opening of the PDU.
8. Remove the clear plastic cover from the power
cable terminal unit.
9. Connect the power cables to the terminals of
the power cable terminal unit.
14. Connect the chassis ground cable of the PDU
to the cabinet frame.
15. Reinstall the rear plate.
16. Connect the power cables of the instruments
and the hour meter to the PDU.
17. Close the rear door.
on the power requirements.
of
Operation Manual
System Installation 4-7