Installing Additional Pin Boards - HP 4062UX Installation Manual

Semiconductor process control system semiconductor parametric test system
Table of Contents

Advertisement

Installing Additional Pin Boards

To install additional HP 16320B Pin Boards in your SWM, refer to Figure 3-1 and perform
the following procedure:
Install pin boards at the installation site only. Wear rubber gloves and a lint
Caution
free cloth cap and gown. Oil, perspiration, hair, brous dust, and dirt degrade
board insulation and reduce measurement accuracy.
1. Remove the protective vinyl cover from the SWM.
2. Remove the protective plate covering the measurement pins.
3. Remove the bottom panel screws, then the bottom panel of the SWM.
4. Remove the required board struts and pin board control cables.
When removing pin board control cables, disconnect each connector from each
Caution
pin board separately. Do
peeling o
5. Remove additional pin boards from their antistatic envelopes.
Wear rubber gloves when handling pin boards. If rubber gloves are not
Caution
available, hold the board by its edges as shown in Figure 3-2.
3-2 Switching Matrix Installation
remove a cable by disconnecting one end and
not
the cable all at once.
Figure 3-1. Accessing Pin Boards

Advertisement