Getting Started
Setting Up a Pulse Signal
Programming
If you want to include this pulse signal in your HP-IB program, use the
Example
following command lines. The comment lines starting with a # are not
required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay
# Pulse stream operating mode is required, but as we start
# from a default status it is not necessary to send a command for
# setting the instrument into pulse stream operating mode.
# Set the period to 50 ns, the pulse width to 20 ns, the leading
# edge to 6 ns and the trailing edge to 10 ns.
:PULSe:PERiod
:PULSe:WIDTh
:PULSe:TRANsition
:PULSe:TRANsition:TRAiling:AUTO
:PULSe:TRANsition:TRAiling
# Set the amplitude to 3.3 Volts, the offset to 1.65 Volts.
:VOLTage
:VOLTage:OFFSet
# Enable the output
:OUTPut
38
OFF
50NS
30NS
6NS
3.3V
1.65V
ON
OFF
10NS