Operation modes
2) Auto tuning
(1) Switching to calibration mode
(2) Starting Auto tuning
18
1. While in operation mode, press the UV CHK switch.
2. The intensity setting value (green) and intensity measurement value
value (red) are displayed.
3. Select the CH for UV intensity measurement.
4. While the intensity setting value (green) and intensity measurement
value (red) are displayed, hold down the SET switch.
5. The intensity measurement value (green) is displayed.
6. Use △/▽ to change the intensity setting to the desired value.
7. Press the SET switch.
8. The intensity measurement value (red) is displayed.
9. Press the SET switch.
10. The intensity setting value (green) and intensity measurement value
(red)
are displayed.
11. Press the EMISSION switch to start irradiation and begin calibration.
When calibration is completed, irradiation will automatically stop.
* If the EMISSION switch is pressed during Auto tuning,
irradiation will stop and an error will be generated.