[1] Procedure of System CD
(1) Power on the ultrasound equipment.
(2) Immediately insert System CD into the ultrasound scanner's CD drive when displaying
HITACHI Logo mark.
(3) Displaying ultrasound boot up image and start up the equipment.
Note1: In case of starting up System CD, it cannot be used the optional function (ex.
SCSI printer, DICOM, Network, ODM and etc)
Note2: After starting up the ultrasound software, if displaying dialog of restart up or not
for discovering new hardware, then select "no".
[2] Procedure to install of the new HDD
Note: In case of attached MO drive (optional EZU-MO1), take off the connector USB cable,
and install as followings. After finished the install, the Drive letter of MO setting "D".
Power on the ultrasound equipment.
(1) Immediately insert System CD into the ultrasound scanner's CD drive when displaying
HITACHI Logo mark.
(2) When the start screen appears, depress "Ctrl + Alt + E" key
Note: It is necessary to depress when displaying start screen. If not depress the key;
repeat from (1) by standby switch, as it is finished start up the ultrasound
software.
(3) Explorer of Windows XP Embedded is started and the file directory in System CD is
displayed.
Note: If displaying dialog of restart up or not for new hardware, then select "no".
(4) Open the "Install" folder from the file directory and double click "shortcut of
diskmgmt.exe" (C:/Install/diskmgmt.exe)
Note: As quite new HDD is sleep mode, if appeared the dialog of discovered new HDD,
press repeating Next button and finish at last.