8.6
Inputs and outputs
The hardware functionality of input and output is fixed when ordering the product. The
available configuration is described in the following table:
Channel
1
2
3
4
118
HW configuration
(fixed when ordering)
Current output
Signal output
Signal output
Relay output
Signal input
Signal output
Relay output
Operating Instructions, 03/2012, A5E03361511-01
SW configuration
available to the user
Current (4-20 mA) HART
Current (0/4-20 mA)
Frequency or pulse
Three-stage analog valve
dosing control
Discrete one or two-valve
dosing control
Operational and alarm status
Current (0/4-20 mA)
Frequency or pulse
Redundant frequency or
pulse
Three-stage analog valve
dosing control
Discrete one or two-valve
dosing control
Operational and alarm status
Discrete one or two-valve
dosing control
Operational and alarm status
Dosing control
Totalizer reset
Remote zero adjust
Force or freeze output(s)
Current (0/4-20 mA)
Frequency or pulse
Three-stage analog valve
dosing control
Discrete one or two-valve
dosing control
Operational and alarm status
Discrete one or two-valve
dosing control
Operational and alarm status
SITRANS FC430 with HART