GE Dash Port 2 Service Manual page 68

Table of Contents

Advertisement

Problem
The AC power/communication mechanism
does NOT extend out of the docking
station's base.
5-10
Troubleshooting: Troubleshooting Procedures
Problems and Solutions Table
Cause
The Top Housing FRU may be
defective, or
Plunger spring and pin are missing.
The locking lever pin or spring are either
missing or not installed correctly.
Dash Port 2 Docking Station
2000966-138
Solution
Complete the following tasks to verify the
Top Housing FRU is functioning properly:
1.
"Separate the Top and Bottom
Housings"
on page 6-17.
2. Move the security lever to the far left
position.
3. Verify the top housing mechanisms
move through their full range of motion.
start position
door open
start position
door closed
stop position
stop position
e. If a mechanism is defective,
the Top Housing FRU"
on page 6-18.
See
"Join the Top and Bottom Housings"
on page 6-25.
039A
042A
"Replace
Revision A