E r r o r , C a u s e a n d R e c o v e r y
Error
Main power failure
Fuse failure
Computer failure
Touch-screen failure
Process fails to start
Plate input failure
Plate output failure
Plate transfer failure
Plate process failure
Plate jamming
Applicator failure
Pipettor tip failure
TM
PREVI
Isola User Manual
Cause
Cable disconnected.
Blown fuse.
Power cable disconnected.
Power cable disconnected.
Covers open.
Input
cassette
is
removed
instrument operation.
Less than five plates in input cassettes.
Plates incorrectly positioned or jammed in
the input cassettes.
Plates loaded agar side down in the input
cassettes.
Output
cassette
is
removed
instrument operation.
Plates incorrectly positioned or jammed in
the output cassettes.
Plates overloaded or jammed in the output
cassettes.
Should not occur during normal operation.
Should not occur during normal operation.
Plates are incorrectly positioned during
loading.
Plates of incorrect size are loaded.
Cartridges are empty.
Applicator robot fails to dispose of used
applicator.
Pipettor trays are empty.
Pipettor fails to dispose of a used tip.
Recovery
Reconnect cable.
Replace fuse.
Contact bioMérieux S.A. or your local
bioMérieux representative.
Contact bioMérieux S.A. or your local
bioMérieux representative.
Close covers.
Replace cassettes.
during
Load plates.
Pause the instrument, remove cassette
and reposition plates.
Pause the instrument, remove cassette
and reload plates.
during
Replace cassette.
Pause the instrument, remove cassette
and reposition plates.
Pause the instrument, remove cassette
and incubate plates.
Contact bioMérieux S.A. or your local
bioMérieux representative.
Contact bioMérieux S.A. or your local
bioMérieux representative.
Pause the instrument, remove cassette
and reposition plates.
Load compatible plates.
Pause the instrument if it has not
already paused or stopped.
Replace cartridges see (Loading PREVI
Isola Applicators, page 4-10).
Remove the applicator manually and
reinitialize the instrument.
Pause the instrument if it has not
already paused or stopped.
Replace pipettor trays (see Loading
PREVI Isola Tips, page 4-13) .
Remove the tip manually and reinitialize
the instrument.
Troubleshooting

Error, Cause and Recovery

9-11