To set up the AT868 analog outputs and totalizer/frequency outputs,
refer to Figure A-2 on page A-2 and follow the instructions in this
section.
Output A and Output C are analog outputs that can be configured
independently. To configure these outputs, complete the following
steps:
1. Press [ESC] , [ENTER] , [ESC] in sequence to enter the user program.
2. Press [ ] until PROG appears and press [ENTER] .
3. Press [ ] until GLOBL appears and press [ENTER] .
4. Press [ ] until I/O appears and press [ENTER] .
5. Press [ ] until OPTN appears and press [ENTER] .
6. Press [ ] to select A or C and press [ENTER] .
1. Press [ ] to select the desired output scale and press [ENTER] .
2. Do one of the following:
If you selected OFF , the meter returns to the Global I/O menu.
Go to Accessing the Totalizer/Frequency Outputs on the next
page.
If you are using a 1-channel meter, go to step 4.
If you are using a 2-channel/path meter, go to step 3.
3. At Channel , press [ ] to select the desired channel and press
[ENTER] .
4. Press [ ] to select the desired measurement parameter and press
[ENTER] .
Note: The measurement units that appear in these prompts are those
selected in the
GLOBL-SYSTM
5. At BASE , enter a flow rate value for the low end of the analog
output range by using the arrow keys. Press [
the cursor to the desired location and press
increase or decrease the number. Then press [ENTER] .
menu earlier in this chapter.
]
and [ ] to move
[ ] and [ ]