Introduction; Calibrating And Testing The Analog Outputs; Calibration - GE AquaTrans AT868 Abridged Manual

Panametrics liquid flow ultrasonic transmitter
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

Introduction

Calibrating and Testing
the Analog Outputs

Calibration

Calibration
Follow the instructions in this chapter to calibrate and test the AT868
analog and totalizer/frequency outputs. Refer to Figure A-4 on page
A-4 in Appendix A, Menu Maps, while following the calibration
instructions.
The AT868 includes one built-in analog output per channel with a
resolution of 5.0 μA (0.03% full scale). Typically, Output A is used
for Channel 1 and Output C is used for Channel 2. However, either
output can be configured for use with either channel. Both the low
and high values for the analog outputs must be calibrated. After the
end points are calibrated, the linearity should be tested.
Note: The zero point of the analog output may be set for either 0 mA
or 4 mA. However, the calibration procedure always uses the 4
mA point, as the meter will extrapolate this value to obtain the
0 mA point.
Prepare for calibration by connecting an ammeter in series with the
load on analog output 1 or 2 as shown in Figure 4-1 below. DO NOT
connect the ammeter directly across the terminals. Refer to Figure 1-5
on page 1-16 for the locations of the analog outputs terminal blocks.
Note: Analog outputs 1 and 2 in the wiring diagram correspond to
analog outputs A and C in the AT868 software.
FREQ1
Figure 4-1: Ammeter Connection
September 2004
AOUT1
Load
Ammeter
4-1

Advertisement

Table of Contents
loading

Table of Contents