10D1476 K-MAG MAGNETIC FLOWMETER INSTRUCTION MANUAL
FIGURE 2-7. MOUNTING DIAGRAM FOR 1- 40 mm (1/25 - 1-1/2") SIZES

2.3.4 Basic Mounting Procedure, Standard Coupling

Select the following procedure applicable for the size meter supplied.

2.3.4.1 1 - 40 mm (1/25 - 1-1/2 inch) Meter Sizes

Refer to Figures 2-1 and 2-6 to supplement the following procedure.
1. Install flange adaptor on meter as shown in Figure 2-7. Align the flange adaptor device (supplied
with meter) so that the mounting stud clearance holes in the meter body are not obstructed.
2. Place one of the gaskets inside the housing end ring next to the inlet end of the ceramic spool; so
that the gasket will be between the pipe flange and the meter body.
3. Insert meter in pipeline in accordance with the flow direction arrow. Next while supporting the
meter, insert one, or more, of the mounting studs through the clearance holes in the upstream pipe
flange, so that these studs enter the holes in the meter body. This will help support the meter.
4. Insert the other (downstream) gasket between the meter outlet and the downstream pipe flange,
with the gasket inside the meter's downstream end ring. The mounting studs can then be guided
through the clearance holes in the downstream pipe flange. Similarly, insert the remaining mounting
studs through the clearance holes provided.