GE MobileLink CB-1000 Installation And Troubleshooting Manual page 12

Wireless communications for ecg
Table of Contents

Advertisement

8
Installation Procedures
12. Select the Listen Port tab.
13. Select/enter the following in the Listen Port tab.
a. Enter the Port Number designated for this MAC 5000 in the
Configuration Capture file. If you are using the suggested
naming convention, enter 3001 if this is MAC5K001, enter 3002
if this is MAC5K002, etc.
b. Select/verify the Reconnect after shutdown box is checked.
14. Select the Packets tab.
15. Verify that all the defaults on this tab are entered:
Linelength:
Input timeout:
Delimiters:
MobileLink
2002783-060
1408 bytes
10 ms
[blank]
07A
08A
Revision D