GE T17 Installation Manual page 86

Ultrasonic flow transducer
Hide thumbs Also See for T17:
Table of Contents

Advertisement

Chapter 5. Maintaining the T17 Transducers
5.2 Using the Insertion Mechanism (cont.)
7. Install a new flange gasket on the isolation valve. Then, fasten a "blind" flange onto the isolation valve with
the bolts removed in the previous step (see Figure 92 below).
8. Disconnect the junction box on the end of the transducer by loosening the compression fitting.
9. Repeat the process to remove the other transducer.
78
Bolt
Gasket
"Blind" Flange
Figure 92: Installing a Blind Flange
Isolation Valve
Model T17 Installation Guide