GE FineLINE 35 oligo Operating Instructions Manual page 29

Table of Contents

Advertisement

Assembly
Step
1
2
3
4
5
6
FineLINE 35 oligo column Operating Instructions 28-9649-57 AA
Action
Wet the bottom filter with ethanol.
Insert the bottom filter unit into the
column tube and tighten it with the
supplied wrench (a).
Wet the adapter filter unit and the adapter rod with ethanol.
Move the adapter lid (b) to its lower stop position.
Insert the adapter into the column tube so that the slot in the adapter lid
is positioned directly above "1" on the column top end-piece (c).
To lock the adapter,
turn the adapter lid
clockwise so that the
slot is positioned
directly above "2" on
the top end-piece (d).
Maintenance 5
Disassembly and assembly 5.4
29