System operation
FineLINE 35 oligo column Operating Instructions 28-9649-57 AA
CAUTION
Ensure that the pressure relief valve is correctly adjusted. Otherwise,
the weakest seal in the system will leak when its limit is exceeded.
CAUTION
Incorrectly fitted tubing may loosen, causing a jet of liquid to spray out.
This is especially dangerous if hazardous chemicals are in use. Connect
the tubing by first inserting the tubing fully, then tightening the
connector fingertight. PEEK tubing should be tightened a further 1/4
turn using the supplied spanner or wrench. Do not tighten Teflon tubing
further as this will damage the end of the tubing.
CAUTION
Do not use in a dusty atmosphere or close to spraying water.
WARNING
Operators should have experience of working with pressure vessels.
WARNING
The design pressure must not be exceeded.
See Section 7.1 Specifications‚ on page 35 for pressure limit.
WARNING
No other solvents than those specified in the Operating Instructions
should be used as they could possibly damage the system.
WARNING
Emergency Stop. When the column is fed by an external pump or
system, the Emergency Stop for that system shall be used in case of an
emergency. Refer to the Operating Instructions for the used pump or
system before using the column.
CAUTION
To avoid precipitation keep all reagents at or above 18° C.
Safety instructions 2
Safety precautions 2.1