Dispensing liquid process chemicals
Filling the dis-
pensing system
46
Insert the service cover into the lower brackets of the machine and
press on it.
Lock the service cover.
Acknowledge the message
with the OK button.
Each time a container is filled or replaced, the
tem
 function must be carried out to remove any air from the dis-
pensing system.
A dispensing system must be filled when:
- The corresponding container has been filled or replaced with a new
one.
- The process chemicals have been changed.
A dispensing system can only be primed if there are no load carri-
ers in the wash cabinet.
System
Open the
 menu.
Machine functions
Open the
Fill dispensing system
Open the
Press the button for the corresponding dispensing system, e.g.,
DOS 1
.
The dispensing system will be filled.
Fill container for dispensing system "x"
Fill dispensing sys-
 option.
 option.