Checking mobile units, modules, and inserts

Mobile units, modules, and inserts should be checked daily to make
sure they are functioning correctly. A checklist is supplied with the
machine.
The following points must be checked:
- Are the guide rails in good condition, and are they securely at-
tached to their mobile units?
- Are the lock nuts for the guide rails tightened securely?
- Are the locking caps in the module connectors of the modular sys-
tem mobile units working properly?
- Are all injector nozzles, irrigation sleeves, and hose adapters se-
curely attached to mobile units or inserts?
- Are all injector nozzles, irrigation sleeves, and hose adapters clear
so that wash water can flow through unhindered?
- Are all caps and fasteners securely attached to the irrigation
sleeves?
If spray arms are present:
- Do the spray arms rotate freely?
- Are the spray arm nozzles free from blockages (see "Maintenance"
– "Cleaning the spray arms")?
- Are the magnets present at the end of the spray arms?
- Are the magnets at the end of the spray arms free from metallic ob-
jects?
Mobile units, modules, and inserts must be inspected periodically
together with the machine after 2000 operating hours or at least
once per year, see "Maintenance" – "Periodic checks".