Technical faults without fault number

Problem
The machine display
is dark.
The wash water in
the machine wash
cabinet is not
drained after a pro-
gram cancellation.
Fault messages and instructions
Possible cause and solution
The machine display is not showing any in-
formation. This can be as a result of the fol-
lowing:
– The display is faulty.
– The connecting cable between the display
and the controls is faulty.
– The plug connection is faulty.
Contact Miele Technical Service.
The machine is not switched on.
Press the I power switch.
An external residual current device (RCD)
has switched.
Inform building services.
An external emergency off switch was acti-
vated.
Inform building services.
In the program head of the reprocessing pro-
gram, the
Drainage after disinfection
ter is active.
Manually disinfect wash water and use the
Manual disinfection
from the wash cabinet (see "Operation,
program cancellation, program cancella-
tion with manual disinfection").