Press the button for the required reprocessing program.
The following message appears on the display:
Program test: Skip program block with . Door can be opened at end of
each program block.
Press OK to confirm the message.
The program start screen is displayed.
Press the  button.
The  button is displayed – the start of the  reprocessing program is
delayed.
Before the water intake, the relevant program phases displayed can
be skipped by pressing the  button.
After each program phase, the following message will appear on the
display before the wash water is drained:
Program test: Door can be opened to take sample.
Press OK to confirm the message.
The  and  buttons are enabled.
To take a sample, press the  buttons to open the door.
Once you have taken the sample, press the  button to close the
door.
The following message will appear on the display:
Continue program test? To continue, press OK. To open door, press 
Press OK to confirm the message.
The wash water is drained and the next program phase begins.
If a sample does not need to be taken after a program phase:
Press the  button.
The wash water is drained and the next program phase begins.
You can skip each program phase when it begins by pressing the
 button.