Appendix A: Interfaces

The modular ergometer system eBike has several interfaces. Digital as well as analog data can be
exchanged with the ergometer. These interfaces thus allow the ergometer to be fully controlled from a
PC or EKG unit and they also permit a mere transfer of data from the ergometer to the EKG unit and
PC.
Analog interface
PORT 2 (9-pin SUB-D)
PORT 3 (5-pin DIN socket)
PORT 1 (9-pin SUB-D)
multi-purpose port
A.1

Digital Interfaces PORT 1 and PORT 3

A.1 a Setup for: Digital ECG Unit
PORT 1:
Digital interface RS 232 + remote start
Pin configuration
(view of connector):
Pin 1
Pin 2 Receive (Input)
Pin 3 Transmit (Output)
Pin 4
Pin 5
Pin 6
Pin 7
Pin 8 remote start EKG unit
Pin 9
Voltages:
Transmit Output (Pin 3 or Pin 1)
Receive Input (Pin 2 or Pin 5)
Baudrate (adjustable via software menu):
default setting:
available settings:
Protocol:
Format:
Start bit:
Data bit:
Stop bit:
Parity:
Handshake:
2018112-003 Rev J
Appendix A: Interfaces
-
-
GND
-
-
-
4800 Baud
1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
ASCII
1
8
1
None
None
eBike, eBike L, eBike EL
Digital interface RS232
Pin 1
Pin 2
PIN 3
Pin 4
Pin 5
logic low
logic high
> +7 V
> -7 V
> +3 V
> - 3 V
PORT 3:
TxD
GND
-
-
RxD
- 75 -

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ebike lEbike el

Table of Contents