GE eBike Service Manual page 70

Hide thumbs Also See for eBike:
Table of Contents

Advertisement

NIBP calibration: calibration at 200 mmHg
Calibration:
Connect pump and calibrated pressure meter to cuff connection.
Depending on the equipment used, you may need a T-adapter to
connect the pump and the pressure meter to the control terminal.
Using the pump of the pressure meter, apply a pressure of exactly
200 mmHg as displayed.
Initiate the calibration with the softkey Calibrate.
NIBP calibration: Calibration is running
During calibration, you will see the message Calibration is running.
At the same time the pressure Value on the NIBP measure Instru-
ment decreased down to 0 mmHG.
An audio signal indicates the end of the calibration procedure and the
Service menu reappears.
Note: According to the German Medical Device Act (MPG), only
GEMS IT authorized personnel is allowed to calibrate the NIBP system.
Note: eBike L / eBike EL have the cuff connection on the right or left
of the couch.
11. NIBP Test
Use a pump to apply pressure from outside.
The current pressure value will be displayed.
Compare the pressure values indicated on the control terminal and on
the display of the pressure meter.
The max. approved tolerance of the pressure readout is +/- 3mm Hg
(see section Functional Test).
Note:
If the pressure reaches 320 mmHg, the safety valve will be
activated. In this case the safety valve can be closed again
by pressing the button "Close Valve" to continue the NIBP
test
12. Ergometer Type
Appears only when no NIBP (non-invasive pressure) module is installed.
For communication with GE electrocardiographs, select ER900.
- 64 -
D: Service Menu
eBike, eBike L, eBike EL
Pressure
compensation
500 ml
Pressure
compensation
500 ml
E r g o m e t e r T y p e
E R 9 0 0
E R 8 0 0
S e l e c t
2018112-003 Rev J