8. PC Mode „Automatic"
With the setting Automatic „Yes", the ergometer will enter the PC
Mode (digital or analog mode) from any operating mode upon detection
of a valid signal.
With the setting Automatic „No", the ergometer will not automatically
change its operating mode. This means that you must select the PC
Mode (digital or analog mode), if you wish to use it.
Note: If the ergometer is controlled by an external EKG unit (serial or
analog/serial control), always select the setting Automatic "Yes".
9. Saddle Calibration (eBike Basic/Comfort only)
The saddle calibration determines the display range (scale from 1 to
40) of the digital saddle height indication.
Select menu item TOP Position and confirm with softkey ENTER
(with eBike comfort only).
Move saddle to the top position (normally 36 cm above the top edge of
the saddle guide).
Confirm setting with SET.
Select menu item BOTTOM Position and confirm with softkey ENTER
(with eBike comfort only).
Move saddle to the bottom position (normally 6.5 cm above the top
edge of the saddle guide).
Confirm setting with SET.
10.NIBP calibration
*
Using the softkeys, choose menu item NIBP
Calibration in the Service menu and confirm
selection with software key Select.
The NIBP calibration menu appears.
*
Zero Adjustment:
During zero adjustment, no pressure must be applied to
the cuff connection at the control terminal.
Note: An adapter nipple (e.g. with an open short length of tubing)
unlocks the cuff connection at the control terminal.
The display shows the current pressure value,
e.g. Offset at 0 mmHg.
Initiate zero balancing with the softkey Start.
Upon zero calibration, you can calibrate the system at 200 mmHg.
2018112-003 Rev J
D: Service Menu
eBike, eBike L, eBike EL
1.
2.
- 63 -