ABB 266 HART Series Operating Instruction page 62

Pressure transmitters
Hide thumbs Also See for 266 HART Series:
Table of Contents

Advertisement

— Sensor related error messages
Error message Tx LCD message
F120.016
Sensor Invalid
F120.016
Sensor Invalid
F118.017
Sensor Memory Fail Sensor memory corrupted.
F114.000
P-dP Sensor Fail
Static Pressure
F112.001
Sensor Fail
Sensor Temperature
F110.002
Fail
NV Sensor Memory
M028.018
Burn Error
— Configuration related error messages.
Error message Tx LCD message Possible cause
Input Simulation
C088.030
Active
Input Simulation
C088.030
Active
Input Simulation
C088.030
Active
M014.037
Configuration Error
M020.042
Replace Info
M020.042
Replace Info
M020.042
Replace Info
62 OI/266/HART-EN Rev. M | 2600T Series Pressure transmitters
Possible cause
The sensor signal is not being updated correctly as a
result of an electronics failure, sensor error or a poorly
connected sensor cable.
The sensor model/version is no longer compatible with
the connected electronic version.
Mechanical damage to the sensor. Loss of fill fluid from
the cell, ruptured diaphragm, broken sensor.
The circuitry for the sampling of the static pressure has
failed.
The circuitry for the sampling of the temperature has
failed.
Writings to the sensor non-Volatile Memory was not
successful.
The P-dP Value produced in output is
derived by the value simulated in input.
The Static Pressure Value produced in
output is derived by the value simulated in
input.
The Sensor Temperature Value produced
in output is derived by the value simulated
in input.
Refer to the Instruction manual to
understand the possible cause of this
error.
The Electronics or the Sensor have been
changed but the replacement operation
has not been executed.
The Electronics or the Sensor have been
changed and a replacement operation for
a new sensor has to be executed.
The Electronics or the Sensor have been
changed.
The replacement has been enabled but
with a wrong direction (SW 2 = 0).
Suggested action
Check cable connection,
check sensor and if problem
persists, the sensor must be
replaced.
The sensor must be replaced.
The sensor must be replaced.
The sensor must be replaced.
The sensor must be replaced.
The sensor must be replaced.
The sensor should be replaced
as soon as possible.
Suggested action
Use a HART configurator (DTM - Hand held) to place
device back in to normal operating mode (Remove
the input simulation).
Use a HART configurator (DTM - Hand held) to place
device back in to normal operating mode (Remove
the input simulation).
Use a HART configurator (DTM - Hand held) to place
device back in to normal operating mode (Remove
the input simulation).
Use a HART configurator (DTM - Hand held) to
correct the configuration.
The replacement operation must be executed: Move
the SW 1 of the electronics in position 1 = Enable
replace mode -Select the SW 2 the element that has
been changed between new Sensor or new
electronics -Power Cycle the device -Move the SW 1
of the electronics in position 0.
The replacement operation must be executed: Only
the data of the electronics can be copied into the
sensor-Move the SW 1 to Enable replace mode
(1)-Select with the SW 2 to New Sensor (1)-Power
Cycle the device-Move the SW 1 to Disable replace
mode (0).
Change the replacement direction (if possible)-The
SW 1 is already set to Enable replace mode (1)-Select
with the SW 2 to New Sensor (1)-Power Cycle the
device-Move the SW 1 to Disable replace mode (0).
9 Error messages
Tx response
Analog Signal to
Alarm
Analog Signal to
Alarm
Analog Signal to
Alarm
Analog Signal to
Alarm
Analog Signal to
Alarm
Analog Signal to
Alarm
no effect
Tx response
no effect
no effect
no effect
no effect
no effect
no effect
no effect

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents