5 Calibration

This section describes how to calibrate the analyzer once it is
operational. Calibration involves measuring the wet-section's
sensitivity to a solution containing a known hydrazine
concentration. Calibrations are initiated via the Cal prompt
displayed on Operator pages, or via the Calibration Level and
Advanced Level menus or as a Scheduled calibration.
Note.
Calibration menus can be accessed from the
Advanced level only – refer to Section 4.3.1, page 20,
steps 2 to 4) to access the Calibrate level.
If calibration standard solution values need to be
changed from their default values, the values must be
set from the Advanced / Calibrate / Calibration Setup
menu only.
Caution.
Do not calibrate the analyzer until the wet-section and
transmitter are installed and ready for operation – refer
to Section 4, page 19.
Allow at least 1 hour for the wet-section to stabilize
before running a calibration.
OI/AHM550–EN Rev. A

5.1 Performing a calibration for the first time

To perform a quick calibration from an Operator page:
1. Press the
key (below the Cal prompt).
Navigator
Tag 1
4.28
4.28
21.7
21.7
N2H4
The Calibrate page is displayed:
Menu
Calibrate
Exit
2. Press the
key (below the Select prompt).
The Calibrate / Hydrazine 1 (4) page is displayed with all
available wet-sections shown:
Calibrate
Hydrazine 1
Back
Use the
/
keys to select the wet-section to
be calibrated.
3. Press the
key (below the Select prompt).
The Hydrazine 1 (2, 3 or 4) / Start Calibration page is
displayed:
Hydrazine 1
Start Calibration
Exit
10 : 34: 52
2012–04–15
ppb
°C
CAL
Select
Select
Yes
23