3.1.5 Connecting the standard solution 2 (high) and reagent tubing

Connections to the standard solution 2 (high) and reagent containers must be made on site. All other internal wet-section tubing
connections are factory-made.
1. Fill the standard solution and reagent containers with the correct solutions – refer to Section Fig. 8.1, page 53, for solution
details.
Referring to Fig. 3.5:
2. Connect the standard solution 2 (high) tubing QD coupling plug
solution 2 (high) container.
3. Connect the reagent solution tubing QD coupling plug
container.
Standard Solution 2
(High) tank – 400 ml
(0.10 US gall.)
Fig. 3.5 Standard solution 2 (high) and reagent tubing connections
OI/AHM550–EN Rev. A
A
to the mating connector
C
to the mating connector
B
at the base of the standard
D
at the base of the reagent solution
Reagent Solution
tank – 200 ml
(0.05 US gall.)
11