3 Installation

3.1 Siting Requirements

3.1.1 Analyzer

Caution.
Mount in a location free from excessive vibration.
Mount away from harmful vapors and dripping fluids.
Note. Mount the analyzer at eye level to enable an
unrestricted view of the front panel displays and controls.
Maximum Distance
20 m (65.6 ft)
A – Maximum Distance between Analyzer and Sensor
–20 °C
(–4 °F)
Min.
B – Within Temperature Limits
C – Within Environmental Limits
Fig. 3.1 Siting Requirements – Analyzer
IM/4690–EN Rev. A
Turbidity
Sensor
55 °C
(131 °F)
Max.
IP66

3.1.2 Turbidity Sensor

To enable the turbidity sensor to be removed easily for
maintenance, ensure an all-round clearance of 200 mm (7.9 in) –
see Section 3.3, page 8 for the overall dimensions of the sensor.
Note. Ensure the sensor is located at an appropriate height
to ensure ease of access during calibration and cleaning.