8.5 Set Up Serial Interface Page
SERIAL INTERFACE
Baud Rate
9600
4800
2400
1200
9 9
Inst. Identity
Parity
NONE
ODD
EVEN
- - - - -
ELECTRICAL CAL
IM/4690–EN Rev. A
Serial Interface
Press
to advance to next page
or
Press
to advance to next parameter.
Baud Rate
Using the
and
keys, select the baud rate required from 1200, 2400, 4800 or 9600 baud.
Instrument Identity
Using the
and
keys, select a value between 1 and 99 to identify the analyzer on the
network.
Parity
keys, select the parity required, None, Odd or Even.
Using the
and
Advance to
Electrical Calibration
8 Configuration and Electrical Calibration
page – see Section 8.6.3, page 33.