COMMAND 135: WRITE POLYNOMIAL COEFFI-
CIENTS
Command-Specific Response Codes
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
COMMAND 137: WRITE UP/DOWN SCALE MODE
Command-Specific Response Codes
1
Type Code Mismatch
2
Invalid Selection
the selection does not correspond with the
transmitter type
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
COMMAND 138: WRITE STATIC PRESSURE TRIM
POINT
Command-Specific Response Codes
3
Passed Parameter too High
4
Passed Parameter too Low
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
11 Excess of correction attempted
COMMAND 139: RESET FACTORY SENSOR
TRIMMING
Command-Specific Response Code
7
In Write Protect Mode
The Write protect link on the transmitter is "on"
COMMAND 140: RESET PRIMARY VARIABLE BIAS
VALUE
Command-Specific Response Code
7
In Write Protect Mode
The Write protect link on the transmitter is "on"
COMMAND 141: SET PRIMARY VARIABLE TO A
SELECTED VALUE
Command-Specific Response Codes
3
Passed Parameter too High
4
Passed Parameter too Low
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
9
Applied Process too High
10 Applied Process too Low
COMMAND 142 : WRITE LIMIT FOR DOUBLE POLY
Command-Specific Response Codes
3
Passed Parameter too High
4
Passed Parameter too Low
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
COMMAND 238: SET LOWER RANGE VALUE ON
O/P SELECTED VALUE
COMMAND 239: SET UPPER RANGE VALUE ON
O/P SELECTED VALUE
Command-Specific Response Codes
3
Passed Parameter too High
4
Passed Parameter too Low
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
COMMAND 242: WRITE SQUARE ROOT CUT OFF
VALUE
Command-Specific Response Codes
3
Passed Parameter too High
4
Passed Parameter too Low
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
COMMAND 244: WRITE DOUBLE POLYNOMIAL
COEFFICIENTS
Command-Specific Response Codes
5
Too few bytes received
the number of bytes received is less than
expected
7
In Write Protect Mode
the write protect link on the transmitter is "on"
TRANSMITTER SPECIFIC COMMANDS
( for 652/653 S TEMPERATURE Transmitters)
COMMAND 141: WRITE CONFIGURATION DATA
Command-Specific Response Codes
2
Invalid Selection
the selection does not correspond with the
transmitter type
5
Too few bytes received
the number of bytes received is less than
expected
6
Transmitter specific command error
the transmitter generates an unexpected error
- 47 -