Operating
6.2 Operation with potentiometer
The stated values are valid under following conditions:
• The distance of the probe to a metal wall is not smaller than stated here Probe mounting
distances - liquids (Page 17) and Probe mounting distances - solids (Page 20)
• The Sensguard (see Options and accessories (Page 76) ) is not used
• Conductive material is not present
Note
The active shield technology in combination with the length of the probe, resulting in
adequate distance between internal measurement electrode and ground electrode, reduces
the influence of various measured capacitance due to the mounting situation and of
moderate material buildup. As a result, no on site calibration is required for general
applications.
30
SITRANS LCS100 I/O Link Version
Operating Instructions, 05/2023, A5E52382716-AA