3.2

Considerations for SMV 3000 Transmitter,

RTD requirements
Thermocouple
requirements
Table 4
16
Use a two-, three-, or four-wire platinum 100 ohm (Pt100) Resistance
Temperature Detector with rated measurement range limits of –200 to
450 °C (–328 to 842 °F) per DIN 43760 standard ( = 0.00385 / / C)
as the input source for the process temperature PV.
Use one of the thermocouple types listed in Table 4 as the input source for
the process temperature.
Thermocouple Types for Process Temperature Sensor
Type
C
E
0 to 1000
J
0 to 1200
K
–100 to 1250
T
–100 to 400
SMV 3000 Transmitter User's Manual
Rated Range Limits
F
32 to 1832
32 to 2192
–148 to 2282
–148 to 752
Continued
Standard
IEC584.1
IEC584.1
IEC584.1
IEC584.1
1/99