Aw Workstation Option; Centricity Ca1000 Option; Injectors; Fluoro Ups (Option) - GE Innova 2121-IQ Biplane Preinstallation Manual

Cardiovascular imaging system
Table of Contents

Advertisement

GE Healthcare

1.8 AW workstation option

AW workstation option is composed of a workstation, 1 or 2 monitors 19'' flat panel in the Control
Room and 1 monitor (flat panel) fixed on suspension (option) and 1 video switcher. Check Innova
Installation Service Manual (JobCard
Splitter) for switcher installation.

1.9 CENTRICITY CA1000 option

Centricity Cardiology CA 1000 V2.0 Preinstallation Guide DOC0241470 in the OEMs of
Refer to :
the Innova 2121-IQ/3131-IQ service manual.

1.10 Injectors

The recommended injectors are:
ILLUMINA (pedestal),
ACIST E2000 Voyager (pedestal)
ACIST CVI (Pedestal and Table mount)
Mark V ProVis (pedestal or rack mounted)
MEDRAD AVANTA (Pedestal and Table mount)
Other injectors (remote or pedestal) require S.O.I..

1.11 Fluoro UPS (Option)

The Innova LC/LP system can be protected with an optional Fluoro UPS.
There are two types of Fluoro UPS:
UL for North America and other 480V – 60Hz countries.
CE for Asia and Europe
32
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
WARNING
DO NOT CONNECT THE UPS BATTERIES PRIOR THE COMMISSIONING OF
THE UPS (INITIAL POWER UP).
IST0093 - Advantage Window Installation (AW and Video
Direction 5177951-4-1EN, Revision 1
1 Description of the System

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3131-iq biplane

Table of Contents