7 X-Ray ON Lamp Distribution
The fitter shall install an X-Ray On lamp in the Exam room, and in the control room if necessary,
outside the patient area, so that they are connected to the equipment through connection terminals.
NOTE:
Innova 2121-IQ/3131-IQ provides 24 VAC for driving X-Ray ON line.
Hospital or contractor to provide 1.5 mm 2 (AWG 14) wires to connect to the system.
X-Ray ON lamp is active during X-Ray exposures.
In the control room, an additional X-Ray on light must be installed if the console (VCIM) indicator
cannot be perceived by all the persons in the control room.
Chapter 6 Electrical Connections
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
WARNING
THE X-RAY ON LAMP MUST BE INSTALLED IN THE EXAM ROOM TO
CONFORM TO INTERVENTIONAL STANDARD IEC/EN 60601-2-43.
SIGNAL INDICATING THE X-RAY ON SHALL BE PERCEPTIBLE BY THE
OPERATOR IN ALL THE LOCATIONS DEFINED FOR THE PERSONNEL WHO
MAY RECEIVE SCATTERED RADIATION.
Illustration 6-12: X-Ray ON cabling between C2 cabinet and PDB
Direction 5177951-4-1EN, Revision 1
167