Calibration Data Sub-Menu - Honeywell SC1000 Owner's Manual

Sensotec sensors signal conditioning, self-calibrating digital indicators
Table of Contents

Advertisement

12.6.5
CALIBRATION DATA
Sub-Menu
or
FULL SCALE VOLT
Menu Item
FULL SCALE CURR
Menu Item
FULL SCALE VALUE
Menu Item
ZERO SCALE VALUE
Menu Item
SHUNT CAL VALUE
Menu Items
KNOWN POINT x/y
This enters the values that will be used by the instrument when it calibrates itself
to the transducer. The menu items that are displayed will change according to
what the
is set for.
CALIBRATION TYPE
The nominal transducer/amplifier output when its full-scale load is applied to it, in
volts or milliamps. This menu item is used by the instrument to set gain of its
amplifier circuit. This value is normally found on the transducer's Certificate of
Calibration or amplifier's Application Note.
The transducer's full-scale output in engineering units.
It is available only with the Shunt Calibration type.
The transducer's zero-scale output in engineering units, which is usually zero.
It is available only with the Shunt Calibration type.
The shunt calibration value must be calculated by using information found on the
transducer's Certificate of Calibration and the following formula:
=
SHUNT CAL VALUE
The terms have the following meanings:
Shunt-Calibration Output: The transducer's output when its shunt calibration
function is activated in Volts or milliamps. Sometimes called "Shunt Cal. Fac-
tor".
Full-Scale Output: The transducer's full-scale output in Volts or milliamps.
Sometimes called "Calibration Factor".
Full-Scale Capacity: The transducer's desired full-scale reading in engineering
units (PSI, lbs, Kg, etc.). You may wish to convert the engineering units of this
value. For example, if the transducer full scale is 3000 PSIG and you wish the
channel to read in bar, you must convert 3000 PSIG into 206.84 bar.
This menu item is only available with the Shunt Calibration type.
To verify proper transducer/amplifier operation, you can activate the transducer's/
amplifier's shunt calibration function while the instrument is in the RUN mode.
Press and hold the [ENTER] button for three seconds; this will activate the trans-
ducer's/amplifier's shunt calibration function and display the reading.
This enters the engineering units for the known-load calibration points. These
points must match the actual loads that you will apply to the instrument during cal-
ibration. The number of calibrations points depends on the
When using the 2-Point Known Load Calibration type, the following menu items
are available:
"
": point 1 of 2, usually 0% of the transducer's full-scale capac-
KNOWN POINT 1/2
ity.
"
": point 2 of 2, usually 100% of the transducer's full-scale
KNOWN POINT 2/2
capacity.
When using the 3-Point Known Load Calibration type, the following menu items
are available:
"
": point 1 of 3, usually 0% of the transducer's full-scale capac-
KNOWN POINT 1/3
Shunt-Calibration Output
-------------------------------------------------------------
Full-Scale Output
SC Series Instruction Manual
×
Full-Scale Capacity
.
CALIBRATION TYPE
page 115