General Specifications
Accuracy of ullage-interface detection
Ullage, interface indication
Tape length
Tape graduation
Tape resolution
Tape accuracy
Diameter of probe (without load)
Minimum detectable tank bottom liquid level
Maximum tank pressure
Accuracy
meets API MPMS Chap 7 2001, table 3 requirements
meets ISO 4268 , IP PMM Part IV
Ambient temperature range
Temperature sensor measurement range
Temperature measurement resolution
Temperature reading
LCD Display
Mechanical coupling
Hazardous environments approvals
IECEx
ATEX
Factory Mutual (FM Approvals)
Multifunctions-Sensor
Ullage detection
Interface detection
Temperature
Innage / Reference height
Tape cleaning device
Tape protection tube
Maintenance
Specifications subject to change without notice.
50461/GTEX MMC/1111
12.

Specifications

±2 mm (± 0.08" approx.)
Audible or visible
15 m/50 ft, 30 m/100 ft, 35 m/115 ft
Metric/English
1 mm / 1/16"
±3.2 mm/30 m (±1/8"/100 ft approx.)
23 mm (29/32" approx.)
4 mm (5/32" approx.)
0,3 bar (4,4 psi)
±0.1°C (0°C to 70°C); ±0.2°F (32°F to 158°F)
-20°C to 50 °C (-4°F to 122°F)
-40°C to 90°C (-40°F to to 194°F)
0.01° or 0.1°, selectable
°C or °F, selectable
8 characters
Q2 (2") or Q1 (1")
Zone 0 Ex ia IIB T4 20°C < Ta < +50°C
II 1 G EEx ia IIB T4 / Tamb 50°C
CL I, DIV 1, GP C&D, T4 Tamb 50°C and
CL I, ZN 0, AEx ia IIB T4 Tamb 50°C
ultrasonic
conductivity
Platinium RTD Pt 1000
additional load (option)
UP / DOWN tape cleaner
on all units equipped with TS storage tubes
modular design / easy exchange of parts