GE Vital Signs enFlow Preventive Maintenance

Iv fluid / blood warming system
Hide thumbs Also See for Vital Signs enFlow:

Advertisement

Quick Links

enFlow
IV Fluid / Blood Warming System
Preventive
Maintenance
Record
Serial No. ________________
Please consult Operator's Manual for Equipment Symbol List
Vital Signs
a GE Healthcare Company

Advertisement

loading

Summary of Contents for GE Vital Signs enFlow

  • Page 1 IV Fluid / Blood Warming System Preventive Maintenance Record Serial No. ________________ Please consult Operator’s Manual for Equipment Symbol List Vital Signs a GE Healthcare Company...
  • Page 2 Employ local regulations to determine the frequency of required testing (i.e. Earthing Impedance, Leakage Current) for the enFlow Warmer (Model 100), Controller (Model 121), and AC Power Pack (Model 120). Frequency Functional and Operational Testing As required by As required by Protocols accrediting body or accrediting body or...
  • Page 3: Enclosure Leakage Current

    2. Plug the cord of the safety analyzer into the Controller/AC Power Pack. 3. Follow steps 5-8 in Procedure A 4. Record the resistance reading. An acceptable reading is a maximum of 100.0 mΩ at a current of 25 A. 2.
  • Page 4 wire is inserted into the tubing just be certain that the wire is in contact with the fluid. 5. Next, put a non-venting cap on the open end of the extension set. 6. Place the cartridge setup in the warmer. 7.
  • Page 5 displayed on the Controller/AC Power Pack front panel. This circumstance obviates the need for the thermometer on this test.) 2. Always run in the normal mode first before switching to the overheat mode. 3. Next, switch to the overheat mode. 4.
  • Page 6 4. The temperatures for this test should be measured within 22.9 cm (9 in.) of both the input and output connections of the Disposable Cartridge. This step is done by inserting T connectors in the direct fluid paths, which will accommodate a temperature probe. Connect the temperature probes to a thermometer capable of measuring between 10 °C and 60 °C with ±...
  • Page 7 © 2010 General Electric Company – All rights reserved. All specifications subject to change without notice. GE, GE Monogram, Vital Signs and enFlow are trademarks of General Electric Company. Enginivity, LLC., a subsidiary of Vital Signs, Inc. Vital Signs, Inc., a General Electric company, doing business as GE Healthcare.
  • Page 8: Enflow Iv Fluid/Blood Warming System Operational Checklist

    enFlow IV Fluid/Blood Warming System Operational Checklist Warmer Serial No__________________ Controller or AC Power Pack Serial No__________________ Warming System Location/Identifier____________________ Date: Test Pass/Fail Date & Pass/Fail Date & Pass/Fail Date & Pass/Fail Date & Initials Initials Initials Initials Recorded Recorded Recorded Recorded Value...