Network Settings - Canon CXDI-60G Service Manual

Table of Contents

Advertisement

2. Installation

7.4 Network Settings

1. Objective
The imaging part communicates with the control computer by using Ethernet [IEEE802.3u
(100Base-TX)] to transfer X-ray images.
The control computer performs DICOM transfer in order to use Ethernet to transfer the obtained
images to the printer and storage device.
This section describes how to set up the TCP/IP that is necessary for the network connection.
Set up the following three items:
1) TCP/IP setting for the control computer network card
2) Network setting for Screwcap.ini
3) Network setting stored in the imaging unit.
2. Preparation
• Keyboard
• Mouse
Connect the keyboard and mouse to the back of the control computer.
Check if the system is connected, and then turn the system on.
3. Setup method
3.1 TCP/IP setting for the control computer network card
Perform the set up by referring to "FC-24VE (or FC-E21A) for CXDI Control Station Service
Manual" -> the chapter "System Manual" -> "Network Setup".
Default values
IP Address:192.168.100.10
SubnetMask:255.255.255.0