Network Settings 1; Objective; Preparation; Setup Method - Canon CXDI-50G Service Manual

Cxdi-1 system
Hide thumbs Also See for CXDI-50G:
Table of Contents

Advertisement

2. Installation Manual

5.4 Network Settings 1

1. Objective

The CXDI-50G imaging part communicates with the control PC by using Ethernet [IEEE802.3u
(100Base-TX)] to transfer X-ray images.
The control PC performs DICOM transfer in order to use Ethernet to transfer the obtained
images to the printer and storage device.
This section describes how to set up the TCP/IP that is necessary for the network connection.
Set up the following three items:
1) TCP/IP setting for the control PC network card
2) Network setting for Screwcap.ini
3) Network setting stored in the CXDI-50G sensor

2. Preparation

• Keyboard
• Mouse
Connect the keyboard and mouse to the back of the control PC.
Check if the system is connected, and then turn the system on.

3. Setup method

3.1 TCP/IP setting for the control PC network card
Perform the set up by referring to "C3S Service Manual" -> the chapter "System Manual" ->
"Network Setup".
Default values
IP Address:192.168.100.10
SubnetMask:255.255.255.0
- 23 -

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents