Notes; Repair And Maintenance; Removing The External Cover; Durability Of Detachable Cable Connector - Canon CXDI-60G Service Manual

Table of Contents

Advertisement

1 Notes

1.1 Repair and Maintenance

In order to ensure safety, the best performance, repair, and maintenance work can only be
performed by the technicians who have received the service training specified by Canon Inc.

1.2 Removing the external cover

When removing the covers (access cover, cover on the power box, etc.,) during maintenance,
repair, etc., perform the work after switching the power off. Never touch the device with wet
hands, as there is a risk of electric shock.
Before proceeding with repairs, ensure that the static accumulated in the bodies of the
installation personnel is discharged. Similarly, before touching the PCBs (when removing them)
or cable connectors, ensure that all static is discharged.

1.3 Durability of detachable cable connector

Cycle life of the detachable cable connector:
Make sure that the power is off before connecting and disconnecting the detachable cable.
Replacing the Sensor Cable Periodically
Communication error may occur if the connection cycle is over 4,000 times.
Monitor the mate/unmate cycles and status of usage at the hospital and encourage periodic
cable replacement by way of prevention and maintenance.
Frequency
Replacement schedule
(by day)
Replacement schedule
(by month)
4. Repair Guide
4000
10 times per day
Five times a day
400 days
16 months
1
cycles.
Two times a day
800 days
32 months
Number of operating day: 300 days a year
2000 days
80 months