Diagnostic Mode; Entering The Diagnostic Mode - Philips V60 User Manual

Respironics hospital ventilator
Hide thumbs Also See for V60:
Table of Contents

Advertisement

Appendix E. Diagnostic mode
Entering the

diagnostic mode

In the diagnostic mode you select the language of software display, set the
date and time, select pressure units, enable software options, and calibrate the
touchscreen.
NOTE:
The diagnostic mode is primarily for use by authorized service
personnel to download software and perform other diagnostic
procedures.
WARNING:
To prevent possible patient injury, do not enter the diagnostic mode while
a patient is connected to the ventilator. Verify that the patient is
disconnected before proceeding.
Enter the diagnostic mode as follows:
1. Make sure the patient is disconnected and the ventilator is powered
off.
2. Press and hold the Accept button on the top-right front of the
ventilator and turn on the ventilator by pressing the ON/Shutdown key.
The screen displays "Press
Ventilation."
again for Diagnostics or wait for
E-1