Installing The Oxygen Manifold Kit; Verifying Ventilator Operation And Audible Alarm; Configuration And Screen Calibration - Philips V60 User Manual

Respironics hospital ventilator
Hide thumbs Also See for V60:
Table of Contents

Advertisement

Installing the oxygen
manifold kit
Verifying ventilator
operation and audible
alarm
Configuration and
screen calibration
If desired, install the oxygen manifold kit as described in the accompanying
instructions.
Perform the following steps to verify ventilator and audible alarm operation:
1. Assemble and install a patient circuit. (See Chapter 5 for instructions
on installing a patient circuit.)
2. Power on the ventilator and verify that it completes the power-on self-
test.
3. Disconnect the proximal pressure airway pressure line from the
ventilator connector, and verify that the Proximal Pressure Line
Disconnect alarm is annunciated (audio, visual, and flashing alarm
LED).
4. Reconnect the proximal pressure line, and manually reset the alarm.
5. Turn the ventilator off.
6. Remove the patient circuit.
The ventilator is ready to be set up for use as described in Chapter 5.
After completing the setup activities described in Chapter 5, set or check the
ventilator settings for language, units of measure, and time in the diagnostic
mode (see Appendix E). Calibrate the screen as required, referring to
Appendix E