Electromagnetic immunity

Guidance and manufacturer's declaration - electromagnetic immunity
The V60/V60 Plus Ventilator is intended for use in the electromagnetic environment specified below.
The user of the V60/V60 Plus Ventilator should assure that it is used in such an environment.
IEC 60601-1-2
Immunity test
test level
Electrostatic dis-
±8 kV contact
charge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast tran-
±2 kV for power sup-
sient/burst
ply lines.
IEC 61000-4-4
±1 kV for input / out-
put lines
Surge
±1 kV differential
mode
IEC 61000-4-5
±2 kV common
mode
Voltage dips, short
U
0%
T
interruptions and
0%
U
T
voltage variations
U
70%
T
on power supply in-
(50 Hz)/
put lines
30 cycles (60 Hz)
IEC 61000-4-11
0%
U
T
(50 Hz)/
300 cycles (60 Hz)
Power frequency
30 A/m
(50/60 Hz) magnet-
ic field
IEC 61000-4-8
U
NOTE:
is the AC mains voltage prior to application of the test level.
T
Compliance level
±8 kV contact
±15 kV air
±2 kV for power sup-
ply lines.
±1 kV for input / out-
put lines
±1 kV differential
mode
±2 kV common
mode
U
for 0.5 cycles
0%
for 0.5 cycles
T
for 1.0 cycle
0%
U
for 1.0 cycle
T
U
for 25 cycles
70%
for 25 cycles
T
(50 Hz)/
30 cycles (60 Hz)
for 250 cycles
0%
U
for 250 cycles
T
(50 Hz)/
300 cycles (60 Hz)
30 A/m
Regulatory compliance
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Mains power quality should be
that of a typical hospital environ-
ment.
Mains power quality should be
that of a typical hospital environ-
ment.
Mains power quality should be
that of a typical hospital environ-
ment. If the user of the V60/V60
Plus Ventilator requires continued
operation during power mains in-
terruptions, it is recommended
that the V60/V60 Plus Ventilator
be powered from an uninterrupt-
ible power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
hospital environment.
D-3