Required facility needs

NOTE:
GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
The VIVID E9 will function on voltages from 100-240 Volts and 50 or 60 Hz. However, if using 220 volt
power in North America, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Dedicated single branch power outlet of adequate amperage (see
all local and national codes which is located less than 2.5 m (8 ft.) from the unit's proposed location
Door opening is at least 76 cm (30 in) wide
Proposed location for unit is at least 0.3 m (1 ft.) from the wall for cooling
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
NOTE:
VIVID E9 has two outlets inside the unit, one is for the B/W printer and one spare.
Power outlets for other medical equipment
Power outlets for test equipment within 1 m (3.2 ft.) of unit
Clean and protected space to store probes (in their cases or on a rack)
Material to safely clean probes (done with a plastic container, never metal)