GE VIVID E9 Service Manual page 236

Hide thumbs Also See for VIVID E9:
Table of Contents

Advertisement

GE H
EALTHCARE
D
GA091568, R
IRECTION
5-8-7-7
LEDs on the DRX board - the GDIF status display
Programming status LEDs exist on the left side of the board. They indicate the programming status of
the GDIF FPGA.
GDIF debug LEDs exist near the lower left side of the board. They are used for GDIF status display.
Figure 5-32 GDIF status display
INT
DONE
GDIF LED 0
GDIF LED 1
GDIF LED 2
GDIF LED 3
GDIF LED 4
GDIF LED 5
GDIF LED 6
GDIF LED 7
GDIF LED 8
GDIF LED 9
5-8-7-8
Troubleshooting hints
During power up, the 4 x 4 LEDs (see
page
During scanning the 4 x 4 LEDs will blink: LEDs in first column will turn ON, then the LEDs in the
next column are lit, then the LEDs in the third column and at last the LEDs in the fourth row. Next,
the sequence will repeat.
If the card starts, the voltages are OK.
If it is artifacts in the picture (during scanning), you may try to interchange the position for the DRX
boards, and scan again. If the artifacts moves to the left or to the right, it indicates an error on a
DRX. If the artifact don't move, the problem is elsewhere in the signal chain.
5 - 46
5
EVISION
TOP MGT 1 LINK UP
TOP MGT 2 LINK UP
BOTTOM MGT 1 LINK UP
BOTTOM MGT 2 LINK UP
TOP MGT 3 LINK UP
TOP MGT 3 LINK UP
BOTTOM MGT 3 LINK UP
BOTTOM MGT 4 LINK UP
BOARD LAST IN CHAIN
ERROR CONDITION
5-45) will be stable ON. If they blink at this time, it indicates an error.
Section 5-8 - Front End Processor (FEP)
Figure 5-31 "LEDs for Nathan (beamforming) circuits" on
VIVID E9 S
M