Inserting Sensors
Important! Always make sure your hands are clean and dry before handling sensors.
Place a sensor into a socket, aligning it with the matching keyways and the multi-pin connector at the bottom. Once a
sensor is plugged in, its corresponding green LED flashes once and if it requires warm-up, then warm-up begins. Below
are approximate times for sensor warm-up.
Non-EC sensor: 0 hours (no warm-up required)
Non-biased EC sensor: 6 hours
Biased EC sensor: 24 hours
If the green LED does not blink, the sensor is not inserted properly.
Note: If the sensor is properly seated and you remove it at any time during its warm-up, it must be returned to its socket
within 20 seconds, or the SensorRAE 4R+ will restart the warm-up cycle from the beginning.
Multi-pin Connector
Top
View
Removing Sensors
Always lift sensors out with your fingers, gripping around the sides.
Checking Sensor Status
You can check the status of a sensor being warmed up by pressing the power button. If the green LED next to the
sensor glows, it is fully conditioned and ready for use.
Bottom
View
RAE Systems by Honeywell
San Jose, CA 95134-1708 USA
E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com
Alignment keys
Alignment keyways (6x)
World Headquarters
3775 N. First St.
Phone: 408.952.8200
Fax: 408.952.8480
Sensor inserted
LED blinks once to
indicate sensor has
been inserted correctly
February 2012
P/N T03-4001-000
Rev. A